Publication

Six Months of Hybrid Closed-Loop Versus Manual Insulin Delivery With Fingerprick Blood Glucose Monitoring in Adults With Type 1 Diabetes: A Randomized, Controlled Trial.

Journal : Diabetes care
Authors : McAuley SA, Lee MH, Paldus B, Vogrin S, de Bock MI, Abraham MB, Bach LA, Burt MG, Cohen ND, Colman PG, Davis EA, Hendrieckx C, Holmes-Walker DJ, Kaye J, Keech AC, Kumareswaran K, MacIsaac RJ, McCallum RW, Sims CM, Speight J, Stranks SN, Sundararajan V, Trawley S, Ward GM, Jenkins AJ, Jones TW, O'Neal DN,
10.2337/dc20-1447 : DOI
33055139 : PMID

Objective

To investigate glycemic and psychosocial outcomes with hybrid closed-loop (HCL) versus user-determined insulin dosing with multiple daily injections (MDI) or insulin pump (i.e., standard therapy for most adults with type 1 diabetes).

Research Design And Methods

Adults with type 1 diabetes using MDI or insulin pump without continuous glucose monitoring (CGM) were randomized to 26 weeks of HCL (Medtronic 670G) or continuation of current therapy. The primary outcome was masked CGM time in range (TIR; 70-180 mg/dL) during the final 3 weeks.

Results

Participants were randomized to HCL ( = 61) or control ( = 59). Baseline mean (SD) age was 44.2 (11.7) years, HbA was 7.4% (0.9%) (57 [10] mmol/mol), 53% were women, and 51% used MDI. HCL TIR increased from (baseline) 55% (13%) to (26 weeks) 70% (10%) with the control group unchanged: (baseline) 55% (12%) and (26 weeks) 55% (13%) (difference 15% [95% CI 11, 19]; < 0.0001). For HCL, HbA was lower (median [95% CI] difference -0.4% [-0.6, -0.2]; -4 mmol/mol [-7, -2]; < 0.0001) and diabetes-specific positive well-being was higher (difference 1.2 [95% CI 0.4, 1.9]; < 0.0048) without a deterioration in diabetes distress, perceived sleep quality, or cognition. Seventeen (9 device-related) versus 13 serious adverse events occurred in the HCL and control groups, respectively.

Conclusions

In adults with type 1 diabetes, 26 weeks of HCL improved TIR, HbA, and their sense of satisfaction from managing their diabetes compared with those continuing with user-determined insulin dosing and self-monitoring of blood glucose. For most people living with type 1 diabetes globally, this trial demonstrates that HCL is feasible, acceptable, and advantageous.